Detection Ranges in Alloys for ElvaX Low alloy steel,
high alloy steel and heat resistant alloys with Ni and Fe-Ni base: |
| Elements |
Detection range (mass fraction), % |
| Beginning of range |
End of range |
| Value |
Absolute Error |
Value |
Absolute Error |
| Al |
0.01 |
0.003 |
2 |
0.1 |
| Si |
0.01 |
0.003 |
5 |
0.2 |
| P |
0.05 |
0.003 |
0.1 |
0.01 |
| S |
0.005 |
0.003 |
0.1 |
0.01 |
| Cr |
0.05 |
0.02 |
30 |
0.3 |
| Ni |
0.05 |
0.03 |
30 |
0.3 |
| Cu |
0.01 |
0.005 |
30 |
0.3 |
| Ti |
0.02 |
0.005 |
4 |
0.1 |
| Mo |
0.005 |
0.002 |
15 |
0.2 |
| V |
0.02 |
0.01 |
3 |
0.15 |
| W |
0.01 |
0.02 |
20 |
0.2 |
| Co |
0.02 |
0.01 |
5 |
0.2 |
| Mn |
0.05 |
0.02 |
20 |
0.2 |
| Nb |
0.005 |
0.005 |
3 |
0.4 |
| Sn |
0.07 |
0.004 |
0.1 |
0.01 |
| Sb |
0.007 |
0.004 |
0.1 |
0.01 |
| Sd |
0.007 |
0.004 |
0.1 |
0.01 |
| Fe |
base |
|
| Comparison of measurement by ElvaX and SRM 1155 Stainless Steel (Cr18-Ni12-Mo2) AISI316 (NIST): |
| Elements |
Certified |
Measured |
| Si |
0,509 |
0,497 |
| P |
0,02 |
0,021 |
| S |
0,0175 |
0,016 |
| Cr |
18,37 |
18,32 |
| Mn |
1,619 |
1,59 |
| Co |
0,105 |
0,097 |
| Ni |
12,35 |
12,38 |
| Cu |
0,173 |
0,169 |
| As |
0,0107 |
0,169 |
| Mo |
2,386 |
2,378 |
| W |
0,11 |
0,108 |
| W |
0,0 |
0,02 |
|